23. A microorganism when viewed under a compound microscope with objective of 10× magnification measured 400 nm in length.
The same microorganism when viewed under a dissecting microscope with 10× lens would measure:

A. 100 m

B. 10 m

C. 40 m

D. 400 nm

Correct Answer: C. 40 μm

A compound microscope with 10× objective typically delivers total magnification of ~100× (10× objective × 10× eyepiece), so the true size of the 400 nm microorganism is 400 nm / 100× = 4 nm actual length. Under a dissecting microscope’s 10× magnification, the image appears 10× larger than actual, measuring 4 nm × 10 = 40 μm.

Detailed Calculation

Step 1: True size from compound microscope
Image size = 400 nm at 10× objective. Standard compound scopes use 10× eyepiece, giving total M = 10× × 10× = 100×.
True length = Image size / Total magnification = 400 nm / 100 = 4 nm.

Step 2: Dissecting microscope measurement
Dissecting (stereo) microscopes provide lower magnification (typically 10×–40× total from 10× objective + eyepiece). At stated “10× lens,” image size = True size × 10 = 4 nm × 10 = 40 × 10⁻⁹ m = 40 μm.

Option Explanations

A. 100 μm: Wrong—assumes ~40× dissecting magnification (4 nm × 25?), overestimating stereo scope power.
B. 10 μm: Wrong—assumes 2.5× magnification, below typical dissecting minimums.
C. 40 μm: Correct—as calculated above for standard 10× stereo optics.
D. 400 nm: Wrong—ignores magnification difference; same image size violates scope comparison.

Microscope Magnification Basics

Microorganism size calculation between compound microscope 10× objective (400 nm image) and dissecting microscope 10× requires understanding total magnification differences—compound ~100× vs stereo ~10×—yielding true size 4 nm and 40 μm image. Essential for GATE Life Sciences optics MCQs.

Magnification Comparison Table

Microscope Type Typical Total Mag. 400 nm Image → True Size 10× Lens Image Size
Compound (10× obj.) 100× (10× eyepiece) 4 nm N/A
Dissecting (Stereo) 10× 4 nm 40 μm 

Exam Formula Application

True size = Measured image / Total M. Dissecting scopes view larger fields at lower power for 3D stereo, unlike compound’s high-res internal focus. Master for microbiology measurement questions.

Leave a Reply

Your email address will not be published. Required fields are marked *

Latest Courses